Payment Terms | L/C, D/A, D/P, T/T, Western Union, MoneyGram |
Supply Ability | 100000 boxes per month |
Delivery Time | 10 working days |
Name | Toxo IgM/Toxo IgG/Rubella IgM/Rubella IgG/CMV IgG/CMV IgM/ Calibrator |
Analyzer | CIA series & POCT in ZECEN |
Storage | 2-8℃ |
Period of validity | 18 months |
Brand Name | ZECEN |
Model Number | Calibrator |
Certification | CE |
Place of Origin | Jiangsu, China |
View Detail Information
Explore similar products
Inflammation Reagent Calibrator PCT CRP IL-6 SAA Reagent Kit Six-Points Accuracy
Kidney Function Calibrator NGAL β2-MG Cys-C RBP H-ALB Reagent Item Six-points
Fertility Hormone Calibrator E2 PROG TESTO Total-hCGβ F-hCGβ PRL FSH LH FE3 AMH
CLIA Calibrator Kit 6-points for Thyroid TSH/TT3/TT4/FT3/FT4/TGAB/TG/TPOAB/TRAB
Product Specification
Payment Terms | L/C, D/A, D/P, T/T, Western Union, MoneyGram | Supply Ability | 100000 boxes per month |
Delivery Time | 10 working days | Name | Toxo IgM/Toxo IgG/Rubella IgM/Rubella IgG/CMV IgG/CMV IgM/ Calibrator |
Analyzer | CIA series & POCT in ZECEN | Storage | 2-8℃ |
Period of validity | 18 months | Brand Name | ZECEN |
Model Number | Calibrator | Certification | CE |
Place of Origin | Jiangsu, China | ||
High Light | CMV IgM Chemiluminescence Calibrator Kit ,CMV IgG Chemiluminescence Calibrator Kit ,Torch toxo igm |
Torch Toxo IgM/Toxo IgG/Rubella IgM/Rubella IgG/CMV IgG/CMV IgM Calibrator for Automatic Immunoassay Analyzer
A calibrator is a reference material, a material or substance whose value for one or more properties is only uniform and well established enough to calibrate a measurement system, evaluate a measurement procedure, or assign a value to a material. There are high and low 2 different concentrations.
Product composition | Two points: Toxo IgM/Toxo IgG/Rubella IgM/Rubella IgG/CMV IgG/CMV IgM |
Six points: Not free | |
Testing method | 1. Treat the calibrator as the sample to be tested, and place it in the sample loading area of the instrument after mixing. Note that when mixing Avoid air bubbles. |
2. During the detection, the system detects the luminescence value (RLU value) of each calibration point, and calculates the luminescence value according to the nominal concentration and luminescence value. | |
The calibration curve is automatically fitted by the instrument. In order to ensure the reliability of the calibration curve, the determination of the calibrator should be | |
Make a double. | |
3. The generated calibration curve can be used for 28 days, but it should be re-calibrated after the following conditions: | |
① Change of batch number of supporting testing reagents; | |
Change of batch number of luminescent substrate | |
②The measured value of the quality control product is not within the scope of quality control or the test results are inconsistent with the clinical situation; | |
③System for major repairs or maintenance. | |
1. For reasons such as methodology or antibody specificity, the same sample was tested using reagents from different manufacturers. | |
Testing with different reagents may give different test results, therefore, the results obtained by testing with different reagents should not be directly | |
They are compared with each other in order to avoid false medical interpretations. | |
2. The quality control product can be used as a reference for the reliability of the current experimental results, and its measured value should be within the | |
within the range allowed by the quality control sheet. Test results should be based on reference value ranges and other clinical factors and results | |
Comprehensive judgment, when the test result is close to the upper limit or lower limit of the reference value range, it can be considered | |
samples for confirmation testing. | |
Storage and expiration date | Stored at 2-8°C, the product is valid for 12 months. Reagents should be used within 1 month after opening, and stored at 2-8°C away from direct sunlight. |
Company Details
Business Type:
Manufacturer
Year Established:
2011
Employee Number:
300~500
Ecer Certification:
Active Member
Zecen Biotech CO., LTD., founded in 2011, is a leading Chinese diagnostics manufacturer specializing in in-vitro diagnostics devices and reagents. Zecen focus on the Chemiluminescent immunoassay (CLIA) field. With all products CE marked, our market presence currently spreads from across China to oth... Zecen Biotech CO., LTD., founded in 2011, is a leading Chinese diagnostics manufacturer specializing in in-vitro diagnostics devices and reagents. Zecen focus on the Chemiluminescent immunoassay (CLIA) field. With all products CE marked, our market presence currently spreads from across China to oth...