Payment Terms | L/C, T/T, Western Union |
Supply Ability | 1000pcs |
Delivery Time | 15-20days |
Packaging Details | Carton |
Brand Name | Fluke |
Model Number | 7008/7037/7012/7011 Refrigerated Temperature Calibration Baths F7008 F7037 F7012 F7011F7312 F9114 F9115A F9116A F9117 F9210 F9230 F9260 F7196B LN2 F1594A F1595A F5430 F742A |
Certification | CCCI |
Place of Origin | China |
View Detail Information
Explore similar products
Fluke Ti480 PRO Infrared Camera FTi480
Fluke TiS20+ / TiS20+ MAX Thermal Imaging Camera FTiS20+ / TiS20+
Fluke 922 Airflow Meter/Micromanometer F922
Fluke 985 Particle Counter F985
Product Specification
Payment Terms | L/C, T/T, Western Union | Supply Ability | 1000pcs |
Delivery Time | 15-20days | Packaging Details | Carton |
Brand Name | Fluke | Model Number | 7008/7037/7012/7011 Refrigerated Temperature Calibration Baths F7008 F7037 F7012 F7011F7312 F9114 F9115A F9116A F9117 F9210 F9230 F9260 F7196B LN2 F1594A F1595A F5430 F742A |
Certification | CCCI | Place of Origin | China |
These four baths operate at temperatures as low as –40 °C, and each one is built using CFC-free refrigerants. Fluke Calibration’s proprietary controller design and unique tank construction produce bath stabilities to ±0.001 °C or better. These baths are so stable and uniform that national labs use them for comparison calibrations and fixed-point cell maintenance.
Each bath (except the 7011) is fully automatable with a bath interface package and Fluke Calibration’s MET/TEMP II automation software package. When we automate a bath, we automate it completely with computer-controlled solenoid valves for precision balancing of the heating and cooling system. MET/TEMP II performs all calibration tasks automatically, using your PC.
With a Fluke Calibration cold bath, you can forget external coolants. Internal refrigeration systems are all that’s needed to reach each bath’s coldest temperature. Most cold baths may be ordered with an optional pumping lid for supplying external cooling requirements.
Each bath has unique characteristics that make it perfect for specific jobs. Some baths are excellent for SPRTs, some are great with thermistors, and some are perfect for maintaining triple point of water cells. A 7008IR bath can even be used to maintain the temperature of a blackbody cone.
Regardless of your application, Fluke Calibration has a bath that gets the job done, and done better than anyone else can do it. Call us today and tell us about your application.
Specifications | ||||
Model | 7008 | 7037 | 7012 | 7011 |
Range | –5 °C to 110 °C | –40 °C to 110 °C | –10 °C to 110 °C | |
Stability | ±0.0007 °C at 25 °C (water) ±0.001 °C at 25 °C (mineral oil) |
±0.002 °C at –40 °C (ethanol) ±0.0015 °C at 25 °C (water) ±0.003 °C at 100 °C (oil 5012) |
±0.0008 °C at 0 °C (ethanol) ±0.0008 °C at 25 °C (water) ±0.003 °C at 100 °C (oil 5012) |
|
Uniformity | ±0.003 °C at 25 °C (water) ±0.004 °C at 25 °C (mineral oil) |
±0.004 °C at –40 °C (ethanol) ±0.002 °C at 25 °C (water) ±0.004 °C at 100 °C (oil 5012) |
±0.003 °C at 0 °C (ethanol) ±0.002 °C at 25 °C (water) ±0.004 °C at 100 °C (oil 5012) |
|
Temperature Setting | Digital display with push-button data entry | |||
Set-Point Resolution | 0.002 °C; high-resolution mode, 0.00003 °C | 0.01 °C; high-resolution mode, 0.00007 °C | 0.002 °C; high-resolution mode, 0.00003 °C | |
Display Resolution | 0.01 °C | |||
Digital Setting Accuracy | ±1 °C | |||
Digital Setting Repeatability | ±0.01 °C | ±0.005 °C | ||
Heaters | 500 and 1000 Watts | |||
Access Opening (call for customs) | 324 x 184 mm(12.75 x 7.25 in) | 162 x 292 mm(6.38 x 11.5 in) | 127 x 254 mm(5 x 10 in) | |
Depth | 331 mm (13 in) | 457 mm (18 in) | 305 mm (12 in) | |
Wetted Parts | 304 Stainless Steel | |||
Power | 115 VAC (±10 %), 60 Hz, 14 A or 230 VAC, 50 or 60 Hz, 8 A, specify | 115 VAC (±10 %), 60 Hz, 16 A or 230 VAC (±10 %), 50 or 60 Hz, 9 A (specify voltage and frequency) | 115 VAC (±10 %), 60 Hz, 14 A or 230 VAC (±10 %), 50 Hz, 7 A, specify | |
Volume | 42 liters (11.2 gal) | 42 liters (11.2 gal) | 27 liters (7.2 gal) | |
Weight | 61 kg (135 lb) | 68 kg (150 lb) | 56.7 kg (125 lb) | |
Size | 610 x 775 x 483 mm (24 x 30.5 x 19 in) | 775 x 768 x 483 mm (30.5 x 30.25 x 19 in) | 762 x 686 x 401 mm (30 x 27 x 15.8 in) | 559 x 686 x 401 mm (22 x 27 x 15.8 in) |
Automation | Use the 1586A Super-DAQ to automate temperature sensor calibration |
Company Details
Business Type:
Manufacturer,Importer,Exporter
Year Established:
2011
Total Annual:
100000-1000000
Employee Number:
15~50
Ecer Certification:
Site Member
Our company produces and sells various stainless steel products, and supports customized production of various non-standard products. Stainless steel products are mainly used in fields such as medicine and universities. Our company also can export kinds of stainless steel prod... Our company produces and sells various stainless steel products, and supports customized production of various non-standard products. Stainless steel products are mainly used in fields such as medicine and universities. Our company also can export kinds of stainless steel prod...
Get in touch with us
Leave a Message, we will call you back quickly!