Payment Terms | T/T |
Supply Ability | 5 Sets per Month |
Delivery Time | 10 Working Days |
Packaging Details | Plywood Box |
Dimension | Interior 160 x 100 x 550cm |
Temperature | NSS ACSS 35℃ ± 1℃ |
Safety Device | Over-temperature Protection |
PH Range | CASS: 3.0~3.2 |
Uniformity | ≦2. 0℃ |
Spray Quantum | 1.0~2.0(ml/80cm2/h) |
Brand Name | iqualitrol |
Model Number | HL-160-SS |
Certification | |
Place of Origin |
View Detail Information
Explore similar products
Copper Accelerated Acetic Acid Salt Spray CASS Test Machine With High Performanc
Environmental Climate Salt Spray Test Chamber / Corrosion Testing Apparatus
neutral Salt Spray Test Chamber with push button pannel in light grey color
270L CASS Salt Spray Test Chamber For Electronic / Car Industry
Product Specification
Payment Terms | T/T | Supply Ability | 5 Sets per Month |
Delivery Time | 10 Working Days | Packaging Details | Plywood Box |
Dimension | Interior 160 x 100 x 550cm | Temperature | NSS ACSS 35℃ ± 1℃ |
Safety Device | Over-temperature Protection | PH Range | CASS: 3.0~3.2 |
Uniformity | ≦2. 0℃ | Spray Quantum | 1.0~2.0(ml/80cm2/h) |
Brand Name | iqualitrol | Model Number | HL-160-SS |
Certification | Place of Origin | ||
High Light | salt spray corrosion test chamber ,salt fog test chamber |
Corrosion Resistance of Coating Electronics Neutral Salt
Spray Testing Chamber
Brief Introduction:
SO2 Salt Spray Test (cyclic) can be used to test the relative resistance to corrosion of product samples that are likely to encounter a combined SO2(sulfur dioxide)/salt spray/acid rain environment during their usual service life. Test specimens are placed in an enclosed chamber, and exposed to 1 of 2 possible changing climate cycles. In either case, the exposure to salt spray may be salt water spray or synthetic sea water prepared in accordance with the requirements of the test standard. The most appropriate test cycle and spray solutions are to be agreed between parties.
The first climate cycle comprises a continuous indirect spray of neutral (pH 6.5 to 7.2) salt water/synthetic seawater solution, which falls-out on to the specimens at a rate of 1.0 to 2.0ml/80 cm²/hour. During this spraying, the chamber is dosed with SO2 gas at a rate of 35 cm³/minute/m³ of chamber volume, for 1 hour in every 6 hours of spraying. The entire test cycle is at a constant chamber temperature of +35C. The number of cycle repeats and therefore the test duration is variable.
The second climate cycle comprises 0.5 hours of continuous indirect spray of neutral (pH 6.5 to 7.2) salt water/synthetic seawater solution, which falls-out on to the specimens at a rate of 1.0 to 2.0ml/80 cm²/hour. This is followed by 0.5 hours of dosing with SO2 gas at a rate of 35 cm³/minute/m³ of chamber volume. This is followed by 2.0 hours of high humidity soak. The entire test cycle is at a constant chamber temperature of +35C. The number of cycle repeats and therefore the test duration is variable.
Features:
1, Chamber material is made of P.V.C Sheet, sturdily and will not be out of shape.
2, Test chamber is heated from five directions to ensure the temperature is uniformity distributed.
3, Spray nozzle is specially made, enable to spray the salt water with considerable speed, then falls on the surface of samples.
4, Hot and moist air keeps the humidity in the test chamber conforms testing request.
5, High precision P.I.D. temperature controller limits the temperature difference within the range of ±0.1℃.
Technical Specifications:
Please Note Small Type of 60 and 90 do not meet standard of ASTM B117.
Model | HL-60-SS | HL-90-BS | HL-120-NS | HL-160-NS | HL-200-NS | ||
Model (Touch Screen) | HL-60-CS | HL-90-CS | HL-120-CS | HL-160-CS | HL-200-CS | ||
Interior dimensions (mm) |
W | 600 | 900 | 1200 | 1600 | 2000 | |
D | 450 | 600 | 850 | 1000 | 1200 | ||
H | 400 | 500 | 500 | 550 | 600 | ||
Exterior dimensions (mm) |
W | 1130 | 1460 | 2200 | 2600 | 3000 | |
D | 630 | 910 | 1200 | 1450 | 1650 | ||
H | 1070 | 1280 | 1450 | 1550 | 1600 | ||
Chamber Volume | 108 L | 270 L | 480 L | 800 L | 1440 L | ||
Solution Tank | 15L | 25 L | 40 L | ||||
Power Supply | AC 220V 50HZ | AC 380V 50HZ | |||||
Ambient temp. & humid. range | +5℃ ~ +35℃ / <85%RH | ||||||
Testing Temperature | NSS ACSS 35℃±1℃ / CASS 50℃±1℃ | ||||||
Air Temperature | NSS ACSS 47℃±1℃ /CASS 63℃±1℃ | ||||||
Solution Consistence | NaCl 5% | ||||||
PH Value | NASS ACSS:6.5~7.2 / CASS:3.0~3.2 | ||||||
Temperature fluctuation | ≦±0.5℃ | ||||||
Temperature uniformity | ≦2. 0℃ | ||||||
Temperature deviation | ≦±2.0℃ | ||||||
Spray quantum | 1.0~2.0(ml/80cm2/h) | ||||||
Chamber Material | polyvinyl chloride (PVC) | ||||||
Heater | Corrosion-resistant titanium tube | ||||||
Temperature sensor | Platinum Resistance (PT-100) | ||||||
Safety device | Over-temperature protection , leakage protection |
Note:
1. The above specifications are for reference only, the actual specifications for the main quotation.
2. Custom made is available according to customers’ special requirements.
Company Details
Business Type:
Manufacturer,Exporter
Year Established:
2000
Total Annual:
700000-1000000
Employee Number:
50~75
Ecer Certification:
Site Member
DongGuan Q1-Test Equipment Co., Ltd. was founded in year 2000, located in DongGuan, a famous industrial city. Mainly specialized in manufacturing, developing and marketing various environmental testing equipment like programmable temperature and humidity test chamb... DongGuan Q1-Test Equipment Co., Ltd. was founded in year 2000, located in DongGuan, a famous industrial city. Mainly specialized in manufacturing, developing and marketing various environmental testing equipment like programmable temperature and humidity test chamb...
Get in touch with us
Leave a Message, we will call you back quickly!