Payment Terms | L/C, T/T |
Supply Ability | 60 Sets Per Month |
Delivery Time | 15 Days |
Packaging Details | Standard Exporting Boxed |
Applied standard | IEC 60990 |
Applied figure | Figure 5 |
Applied another standard | IEC 62368-1 and etc. |
Applied another clause | Clause 5.2.2.2 |
Brand Name | Sinuo |
Model Number | SN2210-54A |
Certification | Calibration Certificate (Cost Additional) |
Place of Origin | China |
View Detail Information
Explore similar products
75g Test Punch With 60° Conical Tungsten Carbide Tip IEC60335-2-24 Clause 22.116
IEC 60335-2-23 Skin or Hair Care Appliance Swivel Connection 50 r/min Rotation
IEC 60335-1 Oxygen Bomb Aging Test Chamber 4000cm3 Stainless Steel Air Bomb Tank
SS Electrical Appliance Tester IEC 60335-1 Clause 20.2 Similar To Test Probe B
Product Specification
Payment Terms | L/C, T/T | Supply Ability | 60 Sets Per Month |
Delivery Time | 15 Days | Packaging Details | Standard Exporting Boxed |
Applied standard | IEC 60990 | Applied figure | Figure 5 |
Applied another standard | IEC 62368-1 and etc. | Applied another clause | Clause 5.2.2.2 |
Brand Name | Sinuo | Model Number | SN2210-54A |
Certification | Calibration Certificate (Cost Additional) | Place of Origin | China |
High Light | Current Measuring Circuit Test Equipment ,IEC60990 Current Measuring Circuit Test Equipment |
IEC 60990 Figure 5 Touch Current Measuring Circuit Test Equipment
Product information:
This test network conforms to the measuring network requirements of IEC 60990 figure 5 touch current and the measurement method of protective conductor current,
(Figure 5- Measuring network, touch current weighted for letgo-immobilization), it is required by IEC 62368-1:2018 clause 5.2.2.2, etc.
An electrical energy source class is determined from both the voltage and the current under normal operating conditions, abnormal operating conditions, and single fault conditions.
Preparation:
Before testing, you need to prepare instruments as follow:
Oscilloscope x1
One or ten times the voltage probe x1
Power supply x1 (the test voltage or frequency that can be adjusted or changed to the required voltage or frequency)
Test Method
Notice
1. During the test, it is necessary to pay attention to safety because you can touch some terminals.
2. According to IEC 60990 standard requirements, the power supply must be in isolation, and zero line and ground wire must be connected together. (the simple way is to whether there exists voltage on zero line and ground wire, if voltage surpass 2V, such as 220V power, the different voltage between zero line and ground wire more than 100V, the zero line and ground wire must be connected.)
3. When testing, please pay attention to the polarity of the power supply. The fire line must be connected to the L end of the contact current test conversion switch. If the connection is wrong, there is a danger of electric shock.
4. Connect the oscilloscope, test the sample, then connect the power after b-terminal line, and do not connect the test line when the power is connected.
Company Details
Business Type:
Manufacturer
Total Annual:
800000-1500000
Employee Number:
30~50
Ecer Certification:
Verified Supplier
Sinuo is an innovation manufacturing company, specializing in the research and development and sales of laboratory construction and testing equipment for electrical appliances (covering the construction of energy efficiency and performance laboratories, as well as supplementary waterproof testing la... Sinuo is an innovation manufacturing company, specializing in the research and development and sales of laboratory construction and testing equipment for electrical appliances (covering the construction of energy efficiency and performance laboratories, as well as supplementary waterproof testing la...