China factories

China factory - China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

  • China,Guangzhou ,Guangdong
  • Verified Supplier

Leave a Message

we will call you back quickly!

Submit Requirement
China ZOLL M Series Defibrillator High Voltage Energy Discharge Capacitor Capacity
China ZOLL M Series Defibrillator High Voltage Energy Discharge Capacitor Capacity

  1. China ZOLL M Series Defibrillator High Voltage Energy Discharge Capacitor Capacity
  2. China ZOLL M Series Defibrillator High Voltage Energy Discharge Capacitor Capacity
  3. China ZOLL M Series Defibrillator High Voltage Energy Discharge Capacitor Capacity
  4. China ZOLL M Series Defibrillator High Voltage Energy Discharge Capacitor Capacity
  5. China ZOLL M Series Defibrillator High Voltage Energy Discharge Capacitor Capacity

ZOLL M Series Defibrillator High Voltage Energy Discharge Capacitor Capacity

  1. MOQ: 1 Piece
  2. Price: USD
  3. Get Latest Price
Payment Terms T/T, Western Union, Paypal
Delivery Time 3-5 working days
Packaging Details Standard Export Packing
Brand ZOLL
Product Category Defibrillation
Product Type High Voltage Capacitor
Machine Application M Series
Part Number 9126-0006
Condition Used with Good Condition
Type Number QL232EW115V23A
Note High Voltage
Warranty 90 Days
Size Small
MOQ One Piece
Brand Name ZOLL
Model Number 9126-0006

View Detail Information

Contact Now Ask for best deal
Get Latest Price Request a quote
  1. Product Details
  2. Company Details

Product Specification

Payment Terms T/T, Western Union, Paypal Delivery Time 3-5 working days
Packaging Details Standard Export Packing Brand ZOLL
Product Category Defibrillation Product Type High Voltage Capacitor
Machine Application M Series Part Number 9126-0006
Condition Used with Good Condition Type Number QL232EW115V23A
Note High Voltage Warranty 90 Days
Size Small MOQ One Piece
Brand Name ZOLL Model Number 9126-0006
High Light High Voltage Defibrillator CapacitorZOLL M Series Defibrillator Capacitor9126-0006 Defibrillator Capacitor

ZOLL M Series Defibrillator Capacitor 9126-0006

ZOLL M Series Defibrillator High Voltage Energy Discharge Capacitor Capacity 9126-0006 QL232EW115V23A

 

Brief Description

ZOLL M Series Defibrillator High Voltage Capacitor 9126-0006

Zoll M Series Defibrillator Capacitor 9126-0006 - This is an important part of the Zoll M series defibrillator. Its main role is to store energy, which can provide the high energy pulse required for the defibrillation operation. The second role is to provide a stable output to ensure the stability of the defibrillation pulse. The third function is rapid discharge, which can complete the discharge process in a very short time and achieve instantaneous high energy output. This rapid discharge property is essential for timely termination of the abnormal rhythm.

 

 

► Removing

Tools Needed

• Number two screwdriver

• Cutting knife

• Polyester tape

 

Step ahead

1. Remove the front panel assembly.

2. Remove the House of Lords.

3. Remove the system board.

4. Remove the battery connection plate.

5. Remove the high voltage charger assembly.

6. Lift up the capacitor assembly from the chassis. (The capacitor assembly is still connected to the high voltage module assembly.)

 

Step 1: Open the high-voltage capacitive plastic isolator and lift the polyester tape.

Step 2: Excess voltage is released using a resistor with a value of about 5 millimetres, 25 watts, for 10-20 seconds. Measure the voltage on the capacitor terminal.

Step 3: Disconnect the inductor and capacitor terminal wires when the capacitor voltage is 0 VDC.

 

 

► More Details

ZOLL M Series Defibrillator Capacitor 9126-0006 sometimes needs to be replaced, the following are some situations that need to be replaced:

1. Press the charge button, can not charge;
2. The capacitor voltage is too high to meet the selected energy;
3. Capacitor voltage > absolute rated maximum value;
4. Unable to charge defib cap;
5. Defibrillators charge too slowly;
6. 4 defibrillators malfunction detected within 20 seconds;
7. Discharge switch in undefined state;
7. Discharge switch in undefined state;
9. The VMON voltage during charging is less than the target energy;
10. When the defibrillator is charged or ready, the capacitor voltage of the defibrillator is greater than the selected energy;
11. The voltage of the defibrillation capacitor has exceeded the absolute maximum acceptable voltage;
12. Defibrillation capacitor voltage does not decrease;
13. The flyback pulse width control circuit is not properly controlled by the processor and gate array.

If you have these problems with your Zoll M Series defibrillator, there may be a problem with the high voltage capacitor, which can be solved by trying to replace the high voltage capacitor.

 

 

 

 

 

 

 

 

 

Thanks for your interest in ZOLL M Series Defibrillator High Voltage Energy Discharge Capacitor Capacity 9126-0006 QL232EW115V23A, please feel free to contact me. Thank you for your time and wish you have a happy life.

Company Details

Bronze Gleitlager

,

Bronze Sleeve Bushings

 and 

Graphite Plugged Bushings

 from Quality China Factory
  • Business Type:

    Manufacturer,Distributor/Wholesaler,Agent,Importer,Exporter,Trading Company,Seller

  • Year Established:

    2015

  • Employee Number:

    20~50

  • Ecer Certification:

    Verified Supplier

China World Technology Medical Equipment Service Group is a company engaged in medical equipments/repair parts and accessories products trading company. The company was established in September 2005, has many famous enterprises at home and abroad (such as America USCO company) to establish a long-te... China World Technology Medical Equipment Service Group is a company engaged in medical equipments/repair parts and accessories products trading company. The company was established in September 2005, has many famous enterprises at home and abroad (such as America USCO company) to establish a long-te...

+ Read More

Get in touch with us

  • Reach Us
  • China World Technology Medical Equipment Service Group
  • TianHe District,GuangZhou,GuangDong
  • https://www.patients-monitor.com/

Leave a Message, we will call you back quickly!

Email

Check your email

Phone Number

Check your phone number

Requirement Details

Your message must be between 20-3,000 characters!

Submit Requirement