Payment Terms | L/C, D/A, D/P, T/T |
Packaging Details | Carton |
Brand | Shanghai |
Color | Metal and glass |
Category | Bulb |
Wattanty | 90 Days |
Packing weight | 1kg |
size | Uniform size |
Application area | Hospitals, private clinics, etc |
Condition | new |
Estimated Ship Time | 3-5 Business Days |
In stock | Enough stock |
Specification | 0.5x0.5cm |
Used for | fully automatic biochemical analyzer light source bulb |
net weight | 0.5 kg |
Shape | standard |
Brand Name | Shanghai |
Model Number | Excellence Kehua KHB420 12V 20W ZY800 ZY1200 fully automatic biochemical analyzer light source bulb |
Certification | CE/CB/ROSH |
View Detail Information
Explore similar products
Original OSRAM HLX 64610 12V 50W Microscope Slit Bulb 12V 50W
OSRAM HLX 64642 24V 150W Long Life Surgical Shadowless Lamp,Halogen lamp
OSRAM Lamp Cup 64653 HLX ELC 24V250W GX5 New Original ,Die-cast aluminum heat
P/N705-0840 12V 20W Hitachi Biochemical Analyzer 7020 7060 7080 7180 7600 bulb
Product Specification
Payment Terms | L/C, D/A, D/P, T/T | Packaging Details | Carton |
Brand | Shanghai | Color | Metal and glass |
Category | Bulb | Wattanty | 90 Days |
Packing weight | 1kg | size | Uniform size |
Application area | Hospitals, private clinics, etc | Condition | new |
Estimated Ship Time | 3-5 Business Days | In stock | Enough stock |
Specification | 0.5x0.5cm | Used for | fully automatic biochemical analyzer light source bulb |
net weight | 0.5 kg | Shape | standard |
Brand Name | Shanghai | Model Number | Excellence Kehua KHB420 12V 20W ZY800 ZY1200 fully automatic biochemical analyzer light source bulb |
Certification | CE/CB/ROSH | ||
High Light | 12V Biochemical Analyzer Light Source Bulb ,Biochemical Analyzer Light Source Bulb |
Excellence Kehua KHB420 12V 20W ZY800 ZY1200 fully automatic biochemical analyzer light source bulb
The light source bulb is the light source bulb of a fully automatic biochemical analyzer. The general process of light transmission is as follows: the composite light emitted by the bulb (12V, 20W placed inside the heat sink) is focused together by a condenser lens (maximizing the use of light source energy), the incident light is converted into parallel light by a collimator lens, and then enters the grating through a slit (to avoid stray light entering). The grating decomposes the composite light into multiple monochromatic lights, reflects them onto a filter (absorbs each monochromatic light), and then sends them to a detector (a column of light-emitting diodes) for photoelectric conversion. Finally, the signal is transmitted to the display screen through a signal processing board for analysis, calculation, amplification, etc.
If the fully automatic biochemical analyzer around you shows an alarm message of "abnormal light bulb energy", you can immediately think that the light bulb may be old and needs to be replaced. However, considering that the fully automatic biochemical analyzer is a relatively large and precise instrument.
You need to find a light bulb of the same model as the automatic biochemical analyzer you have on hand for purchase and replacement.
Before replacing the light bulb, you need to click on "Maintenance" ->"Replace Light Bulb" to enter the maintenance interface, which displays values for 16 wavelengths. From here, you can see the deviation of the values after replacing the light bulb. Generally, the new device value should be above 1.0, and the old device value can also be above 0.7. After replacing the light bulb using this method, it was found that the values of wavelengths 340 and 578 were incorrect. Wavelength 340 corresponds to a value of 0.36, while wavelength 578 corresponds to a value of 0.5, both of which are lower than the standard value of 0.7 mentioned above. Clicking on the alarm option above still shows an "abnormal energy of the light bulb" alarm. After replacing the light bulb, click on the light bulb maintenance interface to check if the device prompts normal operation. If all displays are normal, it indicates that the bulb replacement is reasonable.
Type: | Shanghai Excellence Kehua KHB420 12V 20W ZY800 ZY1200 fully automatic biochemical analyzer light source bulb |
Warranty period: | 90 Days |
Shipping Date: | 3-5 Working Days |
MODLE: | KHB420 12V 20W ZY800 ZY1200 |
Storage temperature: | -20-70°c |
MOQ: | 1 PCS |
COLOR: | Metal and glass |
Delivery Date: | 3 - 5 Working Days |
Package | Carton |
The company has been professionally operating various brands of original medical monitoring instruments since 2015. The company has many professional senior engineers and has sufficient ability to undertake various medical equipment maintenance work. Covering motherboards, sensors, probes, modules, connectors, shells, etc. for various medical monitoring equipment.
We default to express delivery by DHL,Fedex, We will contact you as soon as possible when sending express delivery. If you have any questions after receiving the express delivery, please feel free to contact us at any time.
Company Details
Business Type:
Manufacturer,Distributor/Wholesaler,Agent,Importer,Exporter,Trading Company,Seller
Year Established:
2015
Employee Number:
20~50
Ecer Certification:
Verified Supplier
China World Technology Medical Equipment Service Group is a company engaged in medical equipments/repair parts and accessories products trading company. The company was established in September 2005, has many famous enterprises at home and abroad (such as America USCO company) to establish a long-te... China World Technology Medical Equipment Service Group is a company engaged in medical equipments/repair parts and accessories products trading company. The company was established in September 2005, has many famous enterprises at home and abroad (such as America USCO company) to establish a long-te...
Get in touch with us
Leave a Message, we will call you back quickly!