China factories

China factory - China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

  • China,Guangzhou ,Guangdong
  • Verified Supplier

Leave a Message

we will call you back quickly!

Submit Requirement
China P/N 6640044 Patient Monitor Accessories MAQUET SERVO I SERVO S Ventilator O2
China P/N 6640044 Patient Monitor Accessories MAQUET SERVO I SERVO S Ventilator O2

  1. China P/N 6640044 Patient Monitor Accessories MAQUET SERVO I SERVO S Ventilator O2
  2. China P/N 6640044 Patient Monitor Accessories MAQUET SERVO I SERVO S Ventilator O2
  3. China P/N 6640044 Patient Monitor Accessories MAQUET SERVO I SERVO S Ventilator O2
  4. China P/N 6640044 Patient Monitor Accessories MAQUET SERVO I SERVO S Ventilator O2
  5. China P/N 6640044 Patient Monitor Accessories MAQUET SERVO I SERVO S Ventilator O2

P/N 6640044 Patient Monitor Accessories MAQUET SERVO I SERVO S Ventilator O2

  1. MOQ: 1pcs
  2. Price:
  3. Get Latest Price
Payment Terms T/T(wire transfer), Western Union, Paypal
Delivery Time 3-5 working days
Packaging Details Carton
Product Name P/N 6640044 MAQUET SERVO I SERVO S Ventilator O2 Cell
Brand Maquet
Part No . 6640044
Condition New
MOQ One piece
Warranty 90 days
Compatible Model SERVO I SERVO S
Payment Method T/T ,Paypal ,Western Union
Package Carton
Brand Name MAQUET
Model Number 6640044

View Detail Information

Contact Now Ask for best deal
Get Latest Price Request a quote
  1. Product Details
  2. Company Details

Product Specification

Payment Terms T/T(wire transfer), Western Union, Paypal Delivery Time 3-5 working days
Packaging Details Carton Product Name P/N 6640044 MAQUET SERVO I SERVO S Ventilator O2 Cell
Brand Maquet Part No . 6640044
Condition New MOQ One piece
Warranty 90 days Compatible Model SERVO I SERVO S
Payment Method T/T ,Paypal ,Western Union Package Carton
Brand Name MAQUET Model Number 6640044
High Light P/N 6640044 Patient Monitor AccessoriesO2 Oxygen Cell Patient Monitor Accessories

P/N 6640044 MAQUET SERVO I SERVO S Ventilator O2 Cell

 

 

The O2 cell is housed in the housing of the suction tube and protected by a bacterial filter.

 

Maintenance, including changing the bacterial filter according to the owner's manual. During "preventive maintenance", the bacterial filter must also be replaced. The output voltage provided by the oxygen cell is proportional to the partial pressure of oxygen in the suction tube. At constant ambient pressure, this output is proportional to the percentage of O2 concentration. In each o2 cell, the output signal will remain at a fairly constant level, typically in the 10-17mV range. The lifespan of cells is affected by O2 concentration. If the concentration (at the cell) is % and the expected cell life is in hours, the following conditions apply at 25°C (77oF): o2Conc. x Expected cell lifetime = 500 000% hours. The O2 battery is automatically calibrated each time a pre-use check is performed (if the O2 is connected to the ventilator). If the ventilator is used continuously for a long time, the measured O2 concentration may drop due to the normal degradation of O2 cells. This will activate a nuisance alarm. NOTE: A pre-use inspection is recommended

Item name MAQUET SERVO I SERVO S Ventilator 6640044 O2 Cell
Brand MAQUET
Model 6640044
Use place Hospital Devices; Medical Care; ICU;
Deliver time 3-5 days
Warranty 90 days

 

O2 cell test

Calibrate and check o2 cell at 21% o2 and 100% o2. Check oxygen cells for wear. Due to the different gas mixtures used in this test, the O2 cells will not be calibrated and checked if one gas is missing. (BRE + MON)

 

 

Recommended action if the test fails

1. Check that the connected air supply pressure (air and oxygen) is within the specified range.

2. Replace the O2 battery.

3. Replace the gas module. Replace one gas module at a time.

4. Replace the PC1771 control.

5. Replace the PC1772 monitoring system.

 

 

Recalibrate the oxygen cell

If the ventilator is used continuously for a long period of time, the measured O2 concentration may decrease due to normal deterioration of the oxygen cell. To avoid nuisance alarms in this situation, the O2 unit can be temporarily adjusted during ventilation.

When the O2 cell adjustment function is activated, the oxygen cell is recalibrated so that the current measured O2 concentration is equal to the O2 concentration set by the user. This temporary adjustment remains in effect until the fan is switched off.

Important: Before using the Servo-i, be sure to perform a preliminary check to make sure the O2 unit is properly calibrated.

 

To recalibrate the O2 cell:

1. Press the fixed key menu.

2. Press the Biomed touchpad.

3. Press the O2 cell adjustment touchpad.

4. Press Yes to tap the pad to perform

Adaptation of the O2 cell.

 

 

If you're interested in P/N 6640044 MAQUET SERVO I SERVO S Ventilator O2/Oxygen Cell,feel free to contact us .Thank you .

Company Details

Bronze Gleitlager

,

Bronze Sleeve Bushings

 and 

Graphite Plugged Bushings

 from Quality China Factory
  • Business Type:

    Manufacturer,Distributor/Wholesaler,Agent,Importer,Exporter,Trading Company,Seller

  • Year Established:

    2015

  • Employee Number:

    20~50

  • Ecer Certification:

    Verified Supplier

China World Technology Medical Equipment Service Group is a company engaged in medical equipments/repair parts and accessories products trading company. The company was established in September 2005, has many famous enterprises at home and abroad (such as America USCO company) to establish a long-te... China World Technology Medical Equipment Service Group is a company engaged in medical equipments/repair parts and accessories products trading company. The company was established in September 2005, has many famous enterprises at home and abroad (such as America USCO company) to establish a long-te...

+ Read More

Get in touch with us

  • Reach Us
  • China World Technology Medical Equipment Service Group
  • TianHe District,GuangZhou,GuangDong
  • https://www.patients-monitor.com/

Leave a Message, we will call you back quickly!

Email

Check your email

Phone Number

Check your phone number

Requirement Details

Your message must be between 20-3,000 characters!

Submit Requirement