Payment Terms | T/T, Western Union,PayPal |
Delivery Time | 3-5 working days |
Product name | GE Datex-Ohmeda |
Type | 4901 |
PN | 1500-3378-000 |
Warranty | 90days |
Use with | GE Datex-Ohmeda 7100 Anaesthesia machine |
Stock | In stock |
Mode of transport | Expressage |
Minimum purchasing quantity | 1unit |
Tidal volume range: | 45 to 1500 mL (Volume Control mode) |
Inspiratory pause adjust: | 5% to 60% of inspiratory time (increments of 5%) |
Tidal volume: | 5 to 1500 mL, 1 mL resolution |
Minute volume: | 0 to 99.9 L/min, 0.1 L/min resolution |
Brand Name | GE Datex-Ohmeda |
Model Number | 1500-3378-000 |
View Detail Information
Explore similar products
Patient Monitor Accessories Goldway UT4000 Monitor Blood Pressure Plate Great
Mindray PM7000 Patient Monitor Accessories Keyboard Silicone 7000-20-24395
Well-chosen Patient Monitor Mainboard Goldway G30 40 60 70 80 UT4000B UT6000
Patient Monitor Repair Parts GE Dash4000 Patient Monitor Flat Cable in ready
Product Specification
Payment Terms | T/T, Western Union,PayPal | Delivery Time | 3-5 working days |
Product name | GE Datex-Ohmeda | Type | 4901 |
PN | 1500-3378-000 | Warranty | 90days |
Use with | GE Datex-Ohmeda 7100 Anaesthesia machine | Stock | In stock |
Mode of transport | Expressage | Minimum purchasing quantity | 1unit |
Tidal volume range: | 45 to 1500 mL (Volume Control mode) | Inspiratory pause adjust: | 5% to 60% of inspiratory time (increments of 5%) |
Tidal volume: | 5 to 1500 mL, 1 mL resolution | Minute volume: | 0 to 99.9 L/min, 0.1 L/min resolution |
Brand Name | GE Datex-Ohmeda | Model Number | 1500-3378-000 |
High Light | 1500-3378-000 Disk Ring Bumpers ,GE Datex Ohmeda Disk Ring Bumpers |
GE Datex-Ohmeda Lot# 4901 Bellows Subassy Adult ABA W Disk Ring Bumpers 1500-3378-000 for Datex-Ohmeda 7100 Anaesthesia machine
► Brief Introduction
The Bellows assembly is the interface between drive gas and the patient circuit in the breathing system. The pressure relief valve (or pop-off valve) in the bellows assembly limits pressure in the patient circuit. Excess fresh gas is discharged through the exhalation valve into the gas scavenging system.
The Bellows Pressure Relief Valve (PRV) is normally closed, maintaining approximately 1.5 cm H2O in the breathing circuit in a no flow condition, enough to keep the bellows inflated. It is piloted closed during inspiration and remains closed until the bellows is refilled during exhalation. If the pressure in the patient circuit exceeds 4 cm H2O, the pop-off valve opens to exhaust excess fresh gas flow at a rate up to 4 L/min
► Inspiratory ValveCalibration
1. Remove the bellows and pressure relief valve (pop-off) from the bellows assembly.
2. Attach a patient circuit tube to the Calibrated Flow Orifice test tool.
3. Connect the open end of the patient circuit tube to the inspiration flow patient connection.
4. Set the Bag/Vent switch to Vent.
5. Establish drive gas for the ventilator.
6. Adjust fresh gas flow so the actual Airway Pressure reads 1.05 cm H2
► Breathing circuit parameters
Compliance: | Bag mode: | 1.82 mL/cm H2O |
Mechanical mode: | Automatically compensates for compression losses within the absorber and bellows assembly | |
Circuit volume: | 2.7 L Vent Mode 1.2 L Bag Mode |
► Calibration setup
1. Remove the ABS breathing system from the machine.
2. Remove the Exhalation Valve.
3. Separate the Bellows Module from the Circuit Module.
4. Install the Circuit Module only.
5. Attach a patient circuit tube to the Calibrated Flow Orifice test tool.
6. Insert the Calibrated Flow Orifice into the manifold (PEEP) port.
7. Connect a pressure sensing tee to the inspiration flow patient connection.
8. Connect the open end of the patient circuit tube to the flow port of the pressure sensing tee.
9. Connect a manometer to the pressure sensing port of the tee connector.
► Quality Guarantee
Our company have one division of labour on the basis of specialization, highly cooperative group, rich technical force of medical machines development. Our expert team provides prompt, dependable solutions for sales and repairs on patient monitor, defibrillator machine, ECG/EKG machine and other accessories or repair parts. We have been always focusing on producing superior products and providing first-class customer service.
What calls for special attention is that there are both brand new and second-hand products for sell. Please remember to carefully distinguish them when placing your orders. Otherwise, it will cause unnecessary troubles. If you have any doubts about the product quality, please don’t hesitate to contact us.
Thank you for your time! If you are interested in EDAN F2 F3 F6 F9 US Transducer 8 ultrasound crystals 1 Mhz pink label cable MS3-109301(D) REF 02.01.109301 MPN 02.01.109301017 (01)06944413805484 (21)H22101060151. Please feel free to contact me, if your are interested in any of products.
Company Details
Business Type:
Manufacturer,Distributor/Wholesaler,Agent,Importer,Exporter,Trading Company,Seller
Year Established:
2015
Employee Number:
20~50
Ecer Certification:
Verified Supplier
China World Technology Medical Equipment Service Group is a company engaged in medical equipments/repair parts and accessories products trading company. The company was established in September 2005, has many famous enterprises at home and abroad (such as America USCO company) to establish a long-te... China World Technology Medical Equipment Service Group is a company engaged in medical equipments/repair parts and accessories products trading company. The company was established in September 2005, has many famous enterprises at home and abroad (such as America USCO company) to establish a long-te...
Get in touch with us
Leave a Message, we will call you back quickly!