Payment Terms | T/T, L/C, D/P, T/T |
Supply Ability | 5 sets / 1 months |
Delivery Time | 6~8weeks |
Packaging Details | Plywood cases |
Structure | All-in-one |
Usage | Energy Meter Calibration, Lab Research |
Ambient Temperature | -10℃ to +40℃ |
Power Supply | 3P4W 3×380V±10%, 50Hz±2Hz |
Ambient Humidity | 35% to 85% R.H. |
Voltage Resolution | 0.001% |
Brand Name | GENY |
Model Number | YC-1893R |
Certification | ISO, CE |
Place of Origin | China |
View Detail Information
Explore similar products
Stationary Meter Test System With 0.01 Class Accuracy And Optional Standard
YC-1893R High-grade Three-phase Energy Meter Calibration Device
6/12/24/48 MTUs Electric Meter Test Bench Three Phase Energy Meter Test Bench
Stationary Meter Test System 24 Positions Energy Meter Tester
Product Specification
Payment Terms | T/T, L/C, D/P, T/T | Supply Ability | 5 sets / 1 months |
Delivery Time | 6~8weeks | Packaging Details | Plywood cases |
Structure | All-in-one | Usage | Energy Meter Calibration, Lab Research |
Ambient Temperature | -10℃ to +40℃ | Power Supply | 3P4W 3×380V±10%, 50Hz±2Hz |
Ambient Humidity | 35% to 85% R.H. | Voltage Resolution | 0.001% |
Brand Name | GENY | Model Number | YC-1893R |
Certification | ISO, CE | Place of Origin | China |
High Light | Voltage Stationary Meter Test System ,Powerful Stationary Meter Test System ,Current Stationary Meter Test System |
Power Supply / Electrical System
Voltage | 240/415 V AC supply (+5% & -10%) or customized |
Frequency | 50/60Hz ±2Hz |
Power capacity | > 3 × 6kVA |
Ambient temperature | -10°C … +40°C |
Relative humidity | 35% … 95% |
System Grounding | Neutral solidly grounded |
Surge protection level | ≥ 10 kV/120 Joules (10/1000µs) |
Power source
Voltage | |
Test voltage output (Phase-Neutral) | 3× (24V – 500V) or customized |
Power of the voltage output | 3×100VA |
Resolution | Better than 0.01% of full scale value of range |
Setting accuracy | Better than 0.05% at the final range value |
Stability | Better than 0.005%/h (integration time 150 s) |
Load regulation from 0-max. load | < 0.01% |
Distortion factor | < 0.3% for linear resistance load |
Harmonics | 2nd – 21st free programmable |
Current | |
Test current output | 3× (1mA – 120A) |
Power of the current output | 3×150VA |
Resolution | Better than 0.01% of full scale value of range |
Setting accuracy | Better than 0.05% at the final range value |
Stability | Better than 0.005%/h (integration time 150 s) |
Load regulation from 0-max. load | < 0.01% |
Distortion factor | < 0.3% for linear resistance load |
Harmonics | 2nd – 21st free programmable |
Phase angle | |
Range | 0 -- 360° |
Resolution | 0.01° |
Setting accuracy | 0.1° |
Frequency | |
Range | 45Hz – 65Hz |
Resolution | 0.01Hz |
Reference Standard
Power supply | |
Voltage | AC220V ±10% |
Frequency | 50Hz ±2Hz |
Max power consumption | 45VA |
Using condition | |
Environment temperature | 10ºC … 30ºC |
Relative humidity | 45% … 75% |
Outside magnetic intensity | ≤ 0.025mT |
Frequency range | 45Hz … 65Hz |
Test current | |
Range | 1mA … 120A |
Error | - ±0.02% (50mA…120A) (relative measurement value) - ±0.05% (1…50mA) (relative range full-scale value) |
Current display range | 1.000mA … 120.000A (1A, 10A, 100A) |
Test voltage | |
Range | 50V … 480V (auto range) |
Error | - ±0.02% (50V…480V) (relative measurement value) - ±0.05% (5V…50V) (relative measurement value) |
Voltage display range | 50.000(V) … 480.000(V) |
Power measure error | |
Active power | +0.02% (relative measurement value) 50V … 480V, 0.05A … 120A |
Reactive power | +0.05% (relative measurement value) 50V … 480V, 0.05A … 120A |
Apparent power | +0.05% (relative measurement value) 50V … 480V, 0.05A … 120A |
Test Power Factor(PF=P/F) Error: 0.05% | |
Display range | -1.00000 … +1.00000 |
Energy constant | CH=109 imp/kWh (1A & 0.1A step) CH=108 imp/kWh (10A step) CH=107 imp/kWh (100A step) |
Measuring modes | - 4 wire active/ reactive/ cross-connect reactive - 3 wire active/ reactive/ cross-connect reactive - 2 wire active/ reactive
|
Energy pulse input | |
Frequency | 0.5kHz |
Voltage | 24V |
Measurement range of input | |
Output | TTL pulse, load > 2kΏ |
Dimensions (W * H * D) | 131mm * 432mm * 470mm |
Weight | 16kgs |
The MTS have features as below
- Supporting parallel communication port to test smart meters through RS232 / Ethernet
- Calibrating terminals on the rack for test bench verification
- 1-phase energy meter for testing
- 3-phase energy meter for testing
- Accuracy Testing of reference standard meter
- Pre-warming
- Accuracy test as per IEC
- Creeping test
- Starting current test
- Repeatability of error test
- Dial test
- Influence quantity test related to accuracy tests (Voltage, Frequency, reverse phase sequence, voltage unbalance)
- Influence of harmonic component in voltage and current circuit as per IEC
- Tamper and Fraud Monitoring
- Sub Harmonics as per IEC standards (IEC 62052-11 / 62053-21/22/23)
- Odd Harmonics as per IEC standards
Company Details
Business Type:
Manufacturer,Exporter,Seller
Year Established:
1987
Total Annual:
2,000,000-10,000,000
Employee Number:
50~100
Ecer Certification:
Verified Supplier
Guangzhou GENY Electric Co. Ltd. is a leading manufacturer of energy meter test systems, whose predecessor is Guangzhou Yangcheng Science & Technology Enterprises Co., Ltd. which was established in 1987. We have been dedicated to the energy meter testing industry for over 30 years, and continuou... Guangzhou GENY Electric Co. Ltd. is a leading manufacturer of energy meter test systems, whose predecessor is Guangzhou Yangcheng Science & Technology Enterprises Co., Ltd. which was established in 1987. We have been dedicated to the energy meter testing industry for over 30 years, and continuou...
Get in touch with us