Payment Terms | T/T, Western Union, L/C |
Supply Ability | 100 sets per month |
Delivery Time | 15 days |
Packaging Details | Oil BDV Test Kit packed in Plywood case |
Power supply | AC220V±10%, 50Hz/60Hz |
Display | LCD |
Test Standard | IEC60156 |
Voltage output | AC 0~80KV |
Rate of voltage rise | 2KV/s±10% |
Weight | 41kg |
Keyword | Transformer Oil Tester Kit |
Precision | 3%RDG+0.3%FS |
Brand Name | AI |
Model Number | BDV-I |
Certification | ISO |
Place of Origin | China |
View Detail Information
Explore similar products
ASTM D6304 Automatic Transformer Oil PPM Moisture Content Tester
Transformer Oil Water Content Tester By Coulometric Karl Fischer Titration
Method C Insulating Oil Oxidation Stability Test Set by IEC 61125
Inhibited Mineral Oils Oxidation Characteristics Tester by ASTM D943
Product Specification
Payment Terms | T/T, Western Union, L/C | Supply Ability | 100 sets per month |
Delivery Time | 15 days | Packaging Details | Oil BDV Test Kit packed in Plywood case |
Power supply | AC220V±10%, 50Hz/60Hz | Display | LCD |
Test Standard | IEC60156 | Voltage output | AC 0~80KV |
Rate of voltage rise | 2KV/s±10% | Weight | 41kg |
Keyword | Transformer Oil Tester Kit | Precision | 3%RDG+0.3%FS |
Brand Name | AI | Model Number | BDV-I |
Certification | ISO | Place of Origin | China |
High Light | transformer oil testing kit ,oil bdv test kit ,IEC156 |
I. Introduction:
Insulating oil in electrical power apparatus serves two primary roles, as an insulator and a coolant. As an insulating liquid, oil’s most important property is a high dielectric strength and as a cooling liquid, a low viscosity is paramount. The principal nemeses of insulating oil are oxidation, contamination (particularly moisture, which tends to lower the dielectric properties of insulating oil), and excessive temperature. These enemies are managed through design, regimented installation processes and prudent operation of the oil-filled power apparatus. At all stages of a power asset’s life, testing must be conducted to evaluate the condition of the insulating oil. In fact, for some assets, it can be argued that the oil is tested more frequently than any other component.
Tests that are concerned with the immediate condition and acceptability of the insulating oil include dielectric breakdown tests, moisture measurement by the Karl Fischer (KF) method, and dielectric dissipation factor.
II. Functions:
Breakdown test of user-defined
Proof test
Calibration interface for customers
Uploading data to PC and making word report automatically.
Result print
III.Specifications:
1 | Output voltage | AC 0~80KV |
2 | Rate of voltage rise | 2KV/s±10% |
3 | Measuring precision | 3%RDG+0.3%FS |
4 | Test times setting | 1-9 |
5 | Mixi time settingng | 0~99 second Waiting time setting: 0-9 minute 59second |
6 | Maximum storage capacity | 100 groups of test data |
7 | Display | 256´128 LCD |
8 | Power supply | AC220V±10%, 50Hz/60Hz |
9 | Working environment | 0-50℃, ≤80%RH |
10 | Test standard | IEC60156 |
Company Details
Business Type:
Manufacturer,Trading Company
Year Established:
2018
Total Annual:
1,000,000-5,000,000
Employee Number:
150~200
Ecer Certification:
Verified Supplier
Advanced Instruments Co.,Limited. is a professional supplier which is committed to offer best testing equipment from China for our clients,and our main products are fire rating testing equipment,petroleum products testing equipment,electrical testing equipment geophysical equipment and general labor... Advanced Instruments Co.,Limited. is a professional supplier which is committed to offer best testing equipment from China for our clients,and our main products are fire rating testing equipment,petroleum products testing equipment,electrical testing equipment geophysical equipment and general labor...
Get in touch with us
Leave a Message, we will call you back quickly!