China Whole Blood Sample POCT HbA1c Analyzer 2 Minutes Fully Automated Haematology
China Whole Blood Sample POCT HbA1c Analyzer 2 Minutes Fully Automated Haematology

  1. China Whole Blood Sample POCT HbA1c Analyzer 2 Minutes Fully Automated Haematology

Whole Blood Sample POCT HbA1c Analyzer 2 Minutes Fully Automated Haematology

  1. MOQ: 1
  2. Price: Negotiation
  3. Get Latest Price
Payment Terms D/A, D/P, T/T
Supply Ability 100/Day
Delivery Time 15-20 working days
Packaging Details Carton
Name POCT HbA1c Analyzer Fully Automated Analyzer For Glycosylated Hemoglobin Hba1c Test
Test Time ≤2 minutes
Sample types Whole blood
Sample Volume 4μL
Dimension 192(L)*128(W)*108(H)mm
Weight 800g
Brand Name LABNOVATION
Model Number LD-100
Certification IFCC, NGSP, CE
Place of Origin Shenzhen, China

View Detail Information

Contact Now Ask for best deal
Get Latest Price Request a quote
  1. Product Details
  2. Company Details

Product Specification

Payment Terms D/A, D/P, T/T Supply Ability 100/Day
Delivery Time 15-20 working days Packaging Details Carton
Name POCT HbA1c Analyzer Fully Automated Analyzer For Glycosylated Hemoglobin Hba1c Test Test Time ≤2 minutes
Sample types Whole blood Sample Volume 4μL
Dimension 192(L)*128(W)*108(H)mm Weight 800g
Brand Name LABNOVATION Model Number LD-100
Certification IFCC, NGSP, CE Place of Origin Shenzhen, China
High Light Whole Blood Sample POCT HbA1c AnalyzerPOCT HbA1c Analyzer 2 MinutesPOCT fully automated haematology analyser

POCT HbA1c Analyzer Fully Automated Analyzer For Glycosylated Hemoglobin Hba1c Test

 

The POCT HbA1c Analyzer Fully Automated Analyzer For Glycosylated Hemoglobin Hba1c Test is a fully automated desktop point-of-care analyzer for measuring glycated hemoglobin (HbA1c).
The Test Cartridge is individually packaged containing the reagents necessary for the determination of hemoglobin A1c one test.

 


Feature

 

Boronate affinity Method: The HbA1c assay has no interference from common Hb variants

Ease of use: One-step operation the instrument is ready for use.

• Less than 3-minute test time

• A compact, lightweight instrument, Compact Size - To Minimize Space Requirements.

• With no need for bulky liquid reagents, LD-100 also free up more space in laboratories, clinics, blood banks, and other situations.

• Traceable to the IFCC Reference Method for HbA1c;Results reporting unit: IFCC units (mmol/mol) and NGSP units (%%).

 


Specification

 

Model LD-100
Method Boronate Affinity
Operating Type Fully automatic
Measurement Time ≤2 minutes
Sample Type Whole blood
Measurement Range 4.0-15%
Sample Volume 4μL
Shelf life of test kit 9 months
Weight ≤800g
Dimensions 192(L)*128(W)*108(H)mm

 


Reliable

IFCC calibrator Normal Abnormal Normal Abnormal
Mean HbA1c% 5.4 9.6 5.3 9.2
CV (%) 2.5% 1.9% 2.4% 1.8%

 


Principle

The analyzer works based on the principle of boric acid affinity fluorescence quenching. The content of HbA1c is quantitatively detected by detecting the amount of fluorescent substance changing with the binding and quenching of HbA1c and fluorescent


Performance Parameters

  • Accuracy: Is tested with the reference material as samples, and the relative deviation of the determination results of the analyzer should be within the range of ±8%.
  • Precision:Samples with a concentration of 4.0%-6.5% are detected, sample coefficient of variation (CV) of the test results is not more than 3%.
  • Linearity:In the range of 4%-16% of HbA1c, the linear correlation coefficientrof the test results should be no less than 0.9900.
  • Stability:Within 8 hours after the instrument is turned on and stable, the relative deviation of the test results is not more than ±3%.
  • Functions of the analyzer:The function of report output in English;  Abnormal alarm function.

Attention

  • Analyzer shall be used under good grounding conditions.
  • A three-phase connector is provided for the power cord of the analyzer. When the power socket has a grounding function, the plug shall be inserted into the socket.
  • The installation location shall be free of the direct sunlight and dust.
  • There shall be a level and stable table top, and there shall be no strong magnetic field and vibration within

Certificate

 

 

 

 

Company Details

Bronze Gleitlager

,

Bronze Sleeve Bushings

 and 

Graphite Plugged Bushings

 from Quality China Factory
  • Business Type:

    Manufacturer

  • Year Established:

    2001

  • Total Annual:

    10000000-50000000

  • Employee Number:

    100~600

  • Ecer Certification:

    Active Member

Labnovation Technologies, Inc. is an original manufacturer established in Shenzhen, China since 2001. We are dedicated in developing and manufacturing IVD instruments and reagents including HbA1c analyzer and reagents, rapid test reader & reagents, hematology reagents, clinical chemistry reagent... Labnovation Technologies, Inc. is an original manufacturer established in Shenzhen, China since 2001. We are dedicated in developing and manufacturing IVD instruments and reagents including HbA1c analyzer and reagents, rapid test reader & reagents, hematology reagents, clinical chemistry reagent...

+ Read More

Get in touch with us

  • Reach Us
  • Labnovation Technologies, Inc.
  • 101 and 5th Floor, Building 1, No. 68,18th Road, Guangming Hi-Tech Park, Tangjia Community, Fenghuang Street, Guangming District, Shenzhen 518107 ,China
  • https://www.labnovationtest.com/

Leave a Message, we will call you back quickly!

Email

Check your email

Phone Number

Check your phone number

Requirement Details

Your message must be between 20-3,000 characters!

Submit Requirement