China factories

China factory - Labnovation Technologies, Inc.

Labnovation Technologies, Inc.

  • China,Shenzhen
  • Active Member

Leave a Message

we will call you back quickly!

Submit Requirement
China Individual Test Cartridge POCT HbA1c Analyzer For Glycosylated Hemoglobin Test
China Individual Test Cartridge POCT HbA1c Analyzer For Glycosylated Hemoglobin Test

  1. China Individual Test Cartridge POCT HbA1c Analyzer For Glycosylated Hemoglobin Test

Individual Test Cartridge POCT HbA1c Analyzer For Glycosylated Hemoglobin Test

  1. MOQ: 1
  2. Price: Negotiation
  3. Get Latest Price
Payment Terms D/A, D/P, T/T
Supply Ability 100/Day
Delivery Time 15-20 working days
Packaging Details Carton
Name Individual Test Cartridge POCT HbA1c Analyzer For Glycosylated Hemoglobin Test
Test Speed ≤2 minutes
Sample Type Whole blood
Sample Volume 4μL
Test Range 4%-15%
CV ≤3%
Brand Name LABNOVATION
Model Number LD-100
Certification IFCC, NGSP, CE
Place of Origin Shenzhen, China

View Detail Information

Contact Now Ask for best deal
Get Latest Price Request a quote
  1. Product Details
  2. Company Details

Product Specification

Payment Terms D/A, D/P, T/T Supply Ability 100/Day
Delivery Time 15-20 working days Packaging Details Carton
Name Individual Test Cartridge POCT HbA1c Analyzer For Glycosylated Hemoglobin Test Test Speed ≤2 minutes
Sample Type Whole blood Sample Volume 4μL
Test Range 4%-15% CV ≤3%
Brand Name LABNOVATION Model Number LD-100
Certification IFCC, NGSP, CE Place of Origin Shenzhen, China
High Light Individual POCT HbA1c AnalyzerCartridge POCT HbA1c Analyzer

Individual Test Cartridge POCT HbA1c Analyzer For Glycosylated Hemoglobin Test

 

The Individual Test Cartridge POCT HbA1c Analyzer For Glycosylated Hemoglobin TestI is a automated desktop point-of-care analyzer for measuring glycated hemoglobin (HbA1c) to diabetes diagosis.

 

The Test Cartridge is individually packaged containing the reagents necessary for the determination of hemoglobin A1c one test.

 


Specification

 

Model LD-100
Method Boronate Affinity
Measurement Time ≤2 minutes
Sample Type Whole blood
Measurement Range 4.0-15%
Sample Volume 4μL
Shelf life of test kit 9 months
Weight ≤800g
Dimensions 192(L)*128(W)*108(H)mm
 

 


Principle

 

The analyzer works based on the principle of boric acid affinity fluorescence quenching. The content of HbA1c is quantitatively detected by detecting the amount of fluorescent substance changing with the binding and quenching of HbA1c and fluorescent substance.

 

 

 

 

The Application of Boronate affinity in Individual Test Cartridge POCT HbA1c Analyzer For Glycosylated Hemoglobin Test.

 

After the magnetic beads are pushed into the kit, the borate in the reagent binds to the fluorophore in the magnetic beads, resulting in a large amount of excited fluorescence

After the magnetic beads are pushed into the kit, the borate binds to the fluorophore, producing a large amount of excited fluorescence.

After the sample is added to the reagent, the borate-fluorophore competes with hemoglobin for binding, resulting in a decrease in fluorescence. Because the binding force of unglycated hemoglobin is much greater than that of glycated hemoglobin, the unglycated hemoglobin first binds to the borate-fluorophore, resulting in a sharp drop in fluorescence. The content of unglycated components can be detected during this process.

Next, the glycosylated hemoglobin is bound to a borate-fluorophore. This binding process is slow and therefore the fluorescence quenching process is slow and curvilinear. The fluorescence quenching process in the detection cycle of the instrument can reflect the content of glycosylated hemoglobin.

 

 

 


Feature

 

Boronate affinity Method: The HbA1c assay has no interference from common Hb variants

Ease of use: One-step operation the instrument is ready for use.

• Less than 3-minute test time

• A compact, lightweight instrument, Compact Size - To Minimize Space Requirements.

• With no need for bulky liquid reagents, LD-100 also free up more space in laboratories, clinics, blood banks, and other situations.

• Traceable to the IFCC Reference Method for HbA1c;Results reporting unit: IFCC units (mmol/mol) and NGSP units (%%).


 

 


Performance Parameters

  • Accuracy: Is tested with the reference material as samples, and the relative deviation of the determination results of the analyzer should be within the range of ±8%.
  • Precision:Samples with a concentration of 4.0%-6.5% are detected, sample coefficient of variation (CV) of the test results is not more than 3%.
  • Linearity:In the range of 4%-15% of HbA1c, the linear correlation coefficientrof the test results should be no less than 0.9900.
  • Stability:Within 8 hours after the instrument is turned on and stable, the relative deviation of the test results is not more than ±3%.
  • Functions of the analyzer:The function of report output in English;  Abnormal alarm function.

 

Certificate

 

 

 

 

Company Details

Bronze Gleitlager

,

Bronze Sleeve Bushings

 and 

Graphite Plugged Bushings

 from Quality China Factory
  • Business Type:

    Manufacturer

  • Year Established:

    2001

  • Total Annual:

    10000000-50000000

  • Employee Number:

    100~600

  • Ecer Certification:

    Active Member

Labnovation Technologies, Inc. is an original manufacturer established in Shenzhen, China since 2001. We are dedicated in developing and manufacturing IVD instruments and reagents including HbA1c analyzer and reagents, rapid test reader & reagents, hematology reagents, clinical chemistry reagent... Labnovation Technologies, Inc. is an original manufacturer established in Shenzhen, China since 2001. We are dedicated in developing and manufacturing IVD instruments and reagents including HbA1c analyzer and reagents, rapid test reader & reagents, hematology reagents, clinical chemistry reagent...

+ Read More

Get in touch with us

  • Reach Us
  • Labnovation Technologies, Inc.
  • 101 and 5th Floor, Building 1, No. 68,18th Road, Guangming Hi-Tech Park, Tangjia Community, Fenghuang Street, Guangming District, Shenzhen 518107 ,China
  • https://www.labnovationtest.com/

Leave a Message, we will call you back quickly!

Email

Check your email

Phone Number

Check your phone number

Requirement Details

Your message must be between 20-3,000 characters!

Submit Requirement