Payment Terms | L/C, D/A, D/P, T/T, Western Union, MoneyGram |
Output Signal | 100±20 μA in air |
Load Resistance | 100Ω |
Temperature Range | -30℃ - 50℃ |
Detection Principle | Electrochemical two-electrode |
Detection Range | 0 - 25%VOL |
Maximum Load Concentration | 30%VOL |
Brand Name | KACISE |
Model Number | KQT-O2 |
Certification | CE |
Place of Origin | CHINA |
View Detail Information
Explore similar products
Highly Sensitive TDLAS200 Laser Methane Sensor for Methane CH4 Detection
TDLAS200 High-Sensitivity Laser Methane Sensor
Compact TDLAS200 Portable Methane Gas Detector with 0-5000ppm Measurement Range
Highly Accurate TDLAS200 Laser Methane Detection Module for Methane CH4
Product Specification
Payment Terms | L/C, D/A, D/P, T/T, Western Union, MoneyGram | Output Signal | 100±20 μA in air |
Load Resistance | 100Ω | Temperature Range | -30℃ - 50℃ |
Detection Principle | Electrochemical two-electrode | Detection Range | 0 - 25%VOL |
Maximum Load Concentration | 30%VOL | Brand Name | KACISE |
Model Number | KQT-O2 | Certification | CE |
Place of Origin | CHINA | ||
High Light | Electrochemical 2 Electrode Gas Detector ,RapidO2 Gas Detector ,Alphasense Gas Detector |
The RapidO2 Gas Detector by KACISE is a cutting-edge solution for real-time oxygen monitoring, designed as a reliable alternative to Alphasense. Featuring fast response times (T90 < 20s) and quick return-to-zero functionality, it is ideal for medical, industrial, and environmental applications. With a detection range of 0-25% VOL and high sensitivity, the RapidO2 ensures accurate and stable performance. KACISE’s innovative engineering ensures low power consumption, long-term stability, and excellent repeatability, making this detector a trusted choice for ensuring safety and air quality in diverse settings.
1.Oxygen alarms in the mining field
2.Medical oxygen generators
3.Air quality detection instruments
4.Air purification equipment in underground parking lots and commercial areas
5.Air quality detection in vehicles, as well as the fresh air systems of in-vehicle purification equipment
1. High precision and long service life
2. Fast response speed and quick return to zero
3. Low power consumption and high sensitivity
4. Wide linear range
5. Strong anti-interference ability
6. Excellent repeatability and stability
Project | Parameter |
Model | KQT-O₂ |
Detection Principle | Electrochemical two-electrode |
Detected Gas | O₂ |
Detection Range | 0 - 25%VOL |
Maximum Load Concentration | 30%VOL |
Output Signal | 100±20 μA in air |
Zero-point Drift | 0.6%VOL |
Resolution | 0.1%VOL |
Response Time | T90 < 20s |
Bias Voltage | 0 |
Load Resistance | 100Ω |
Temperature Range | -30℃ - 50℃ |
Humidity Range | 15% RH to 90% RH (no condensation) |
Repeatability | < ±2% output signal |
Long-term Stability | < 2% signal / month |
Linearity | Linear, regression coefficient R² = 0.999 |
Working Pressure | 90 to 110 kPa |
Shelf Life | 12 months after delivery |
Service Life | 2 years |
Company Details
Business Type:
Manufacturer
Year Established:
2008
Total Annual:
8000000-10000000
Employee Number:
100~150
Ecer Certification:
Verified Supplier
Xi'an Kacise Optronics Co.,Ltd. is a professional manufacturer of measuring instruments , located in China xi’an City. Since 2008, Xi'an Kacise focused on the development and production of pressure, level and flow measuring instruments . We have been specializing... Xi'an Kacise Optronics Co.,Ltd. is a professional manufacturer of measuring instruments , located in China xi’an City. Since 2008, Xi'an Kacise focused on the development and production of pressure, level and flow measuring instruments . We have been specializing...
Get in touch with us
Leave a Message, we will call you back quickly!